External and internal circulating with powerful pumping capacity
CW-G series is designed for controlling temperature in external instrumentation, or for immersion application. Pump pressure is adjustable in five stages and all models come with built-in tab water cooling coil.
CW-05G (5.0L) – Cat.No. JTH-AAH52306U
CW-10G (10.0L) – Cat.No. JTH-AAH52316U
CW-20G (20.0L) – Cat.No. JTH-AAH52326U
CW-30G (30.0L) – Cat.No. JTH-AAA52336U
Operating Features:
- 5°C above room temperature to 150°C
- Uniformity ±0.1°C at 50°C (based on various conditions)
- Microprocessor PID control/ Auto-tuning/ Calibration
- Digital timer: 1 min to 99 hr 59 min, delayed ON/ OFF
- Over-temperature limiter and low fluid level protection
- 3 different temperature values memorable without auto-tuning
- Digital LED display (0.1°C resolution) with splash-proof keypads
- RS-232 interface available
Constructional Features:
- Exclusive variable-speed pump for accurate, stable, and efficient temperature management
- (1 to 3 steps of pump for internal circulation/ 4 and 5 steps for external circulation)
- Powerful pressure and suction pumps provide contrast pressure and flow rate over the full temperature range
- Baffle plate screens circulating pump, heater and sensor to minimize shaking bath fluids and protect sample and operators from direct contact
- Dry-running protection; automated heating element cut-off and audible and visual warning
- Seamless stainless steel tank with a reservoir drain that resists corrosion and make cleaning easy
- Space saving design uses minimal bench space
Specifications:
| Model |
CW-05G |
CW-10G |
CW-20G |
CW-30G |
| Bath volume |
5.0L (0.18cu ft) |
10.0L (0.35cu ft) |
20.0L (0.7cu ft) |
30.0L (1.05cu ft) |
| Temperature |
Range 1) |
Amb. +5°C to 150°C (Amb. +9°F to 302°F) |
| Accuracy |
±0.05°C at 50°C (±0.09°F at 122°F) |
| Uniformity |
±0.1°C at 50°C (±0.18°F at 122°F) |
| Heat up time 2) |
50°C(122°F) within 30 min |
| Pump |
Max. pressure,
max. flow rate |
5.9 PSI (406.8 millibar), 7.4 gallons/minute (28.0 litres/minute) |
| Max. suction flow rate 3) |
2.6 gallons/minute (10.0 litres/minute) |
| Dimensions |
Bath opening/ depth
(W×L/ D) |
150×99/ 160mm |
211×154/ 180mm |
290×214/ 200mm |
300×264/ 230mm |
(5.9x3.9/ 6.3”) |
(8.3x6.1/ 7.1”) |
(11.4x8.4/ 7.9”) |
(11.8x10.4/ 9.0”) |
| Overall (W×L×H) |
220×395×424mm |
281×450×444mm |
360×510×464mm |
370×560×494mm |
(8.7x15.6x16.7”) |
(11.1x17.7x17.5”) |
(14.2x20.1x18.3”) |
(14.6x22.0x19.4”) |
| Net weight |
12.5kg (27.6lb) |
15.0kg (33.0lb) |
17.5kg (38.6lb) |
19.5kg (43.0lb) |
Electrical requirements
(120V, 50Hz) |
50/60Hz, 8.6A |
50/60Hz, 8.6A |
50/60Hz, 8.6A |
50/60Hz, 8.6A |
1) If you operate the unit in 1 to 3 steps of pump, the temperature range can be from Amb. +5°C but in 4 and 5 steps, the range will be from Amb. +15°C.
2) The heat-up time is measured on a 230V electrical condition with 1000W heater power.
3) The additional suction function is optional. |