Single-probe,
"workhorse" liquid handler delivers a robust
performance for fast and accurate direct injection onto
HPLC or LC/MS systems. Suitable for a wide variety of
applications, with the ability to pierce thick septa.
Performs injection, fraction collection and re-injection
of collected fractions on one platform.
Manufacturing Standards |
Meets applicable safety and EMC certification standards; UL and CE certified |
Arm Speed |
>45.7 cm/sec (>18 in./sec) in X dimension
>40.6 cm/sec (>16 in./sec) in Y dimension |
Communication Interface |
RS-232 or GSIOC; Four inputs (contact closure, TTL, or open-collector), four relay outputs and one switched +24V DC 1A output |
Accuracy |
Liquid handling: 99.7% with 50 µL of water |
Front Panel |
8-character display, START and emergency STOP soft keys |
Horizontal Motion Strength |
X: 5 kg (11.1 lbs.)
Y: 7 kg (15.6 lbs.) |
Injection Carryover† |
<0.1% |
Locator Plate Capacity |
Up to five Code 200-Series racks, up to seven Code 20- or 30-Series racks (or a combination of up to five racks of both types), or one Code 500-Series and one Code 200-Series rack |
Maximum Syringe Flow Rates for Water |
Syringe Size
(µL) |
Max. Flow Rate
(mL/min) |
100 |
3.37 |
250 |
8.43 |
500 |
16.87 |
1000 |
33.75 |
5000 |
100 |
10,000 |
100 |
25,000 |
100 |
|
Power Requirements |
Frequency: 50–60 Hz
Voltage: 100–120 or 220–240V, mains voltage fluctuations not to exceed ±10% of the nominal voltage
Current rating: 2.0A for 100–120V or 1.0A for 220–240V |
Probe Positioning Performance |
Accuracy:±0.5 mm in X/Y dimensions±1 mm in Z dimension
Repeatability:±0.25 mm in X/Y/Z dimensions |
Probe Rinse |
Through a dedicated rinse station for rinsing the inside and outside of the probe; selectable rinse volume and flow rate. Optional inserts for level-sensing, nonleveling-sensing and flow rinse |
Sampler Type |
X/Y/Z with stationary rack design |
Software |
Via Gilson UniPoint™ System Software or TRILUTION™ Software |
Syringe Capacity |
100, 250 or 500 µL; 1, 5, 10 or 25 mL |
Syringe Pump |
Internal, integral, high-precision, single-piston |
Vertical Punch Strength |
4.9 kg (11 lbs.) |
Reproducibility† |
Liquid Handling:
1.37% CV with 50 µL of water
Injection:
<0.9% CV with 1-mL syringe and total loop filling method |
Dimensions (w x d x h) |
91.4 x 61 x 55.8* cm (36 x 24 x 22* in.)
*Maximum height. Z-arm height is adjustable to accommodate vessel heights between 1 and 150 mm (dependent on installed Z-arm). |
Weight |
39.9 kg (89 lbs.) |