IR Prestige
IR Prestige | Discover the New Worlds Opened by the IRPrestige

FTIR Special Packages

Shimadzu was the first company to offer dynamic auto-alignment as a standard feature in an FTIR spectrometer in its' price range. With state-of-the-art technology, the Shimadzu dynamic alignment system continuously maintains optimal alignment during data acquisition, ensuring consistently reproducible spectra without the need for tedious mechanical adjustments.

Features

The Near IR/ Mid IR/ Far IR World

The Standard optical bench features a Mid IR measuring range of 7,800 to 350cm-1. Optional kits can provide extended measuring range in the Near IR and Far IR from 12,500-240cm-1.

The Signal-to-Noise World

Unprecedented S/N for an FTIR in its price-class; peak-to-peak S/N is 40,000:1 or better, guaranteed!

The Wide World of Applications

The IRPrestige covers a wide range of applications and can be easily modified with a variety of beam Splitters, detectors, and light sources.

The Advanced Dynamic Alignment World

Optimal interferometer alignment is maintained automatically for continuous optimization and exceptional reproducibility.

The World of Reliability

The software validates instrument performance to ensure compliance. Self Diagnostics and Status Monitoring ensure that the instrument is performing optimally and that accurate data is being obtained.

The World of Simplicity

New FTIR operators will love the Quick-Start Accessory function that recognizes installed accessories and automatically sets appropriate measurement methods.

Software

The IRsolution World

The powerful new IRsolution software increases the ease of use of the IRPrestige. The comprehensive package includes standard data processing functions, spectral search, quantitation and easy report generation. Tailor the software to fit your application with the customizable user interface to suit a range of users.

 

IR Prestige
Interferometer Michelson inferometer (30 degree incident angle)
Advanced dynamic alignment system
Sealed and desiccated inferomter with an automatic dryer (Patent pending)
Optical System Single beam optics
Beam Splitter Germanium-coated KBr plate for Middle IR (Standard)
Germanium-coated Csl plate for Middle/Fair IR (Optional)
Silicon-coated CaF2 plate for Near IR (0ptional)
Light Source Air cooled ceramic for Middle/Far IR (Standard)
Tungsten lamp for Near IR (Optional)
Detector DLATGS detector with temperature control for Middle/Far IR (Standard)
MCT(Hg-Cd-Te) detector with liquid nitrogen cooling for Middle IR(Optional)
InGaAs detector for Near IR(Optional)
Wavenumber range 7,800 cm ~ 350 cm-1
12,500 cm ~ 240 cm-1(Optional)
Resolution 0.5,1,2,4,8,16 cm-1(Middle/Far IR)
2,4,8,16 cm-1(Near IR)
Signal-to-Noise Ratio Greater than 40,000:1
(4 cm-1 resolution, 1 minute accumulation, around 2,100 cm-1, peak-to-peak)
Mirror speed 3-step selecction of 2.8, 5, or 9 mm/sec
Scanning at 4 cm-1 takes from 2-3 sec
Data sampling He-Ne laser with 30 months guarantee
Gain control Automatic or manual from x1 - x128
Sample compartment Automatic accessory recognition
W 200 mm x L 230mm x H 170mm
Dimensions W 600mm x L 680mm x H 290mm
Weight 54kg

 

 

Application Notes

PDF Oil in Water (PDF)
PDF Quantitative Analysis of Moisture in Tobacco (PDF)
PDF Section Analysis of Wheat Grain (PDF)
PDF Sugar - Honey (PDF)

 

  Privacy Policy | Comments