The 215 Direct-Inject Liquid Handler/Injector is designed to handle the demands of smaller injection volumes—delivers microliter volumes with precision and accuracy. Capable of injecting volumes of 500 nL to 5 mL from 384- or 96-well microplates, vials or tubes.

Features

  • Accommodates up to 17 standard- or deep-well microplates
  • Increases throughput with valve switching speeds five times faster than the standard 215 Liquid Handler/Injector
  • Optional Peltier Racks for cooling and heating of samples in standard microplates
  • Ideal for injecting 1 µL (or less) directly into the 841 Injection Module to minimize carryover and maximize precision; can also accommodate large-volume injections without reconfiguring the instrument
  • Add a second 841 Injection Module to increase throughput while preserving bench space

Specifications

Manufacturing Standards

Meets applicable safety and EMC certification standards; UL and CE certified

Arm Speed

X: >45.7 cm/sec (>18 in./sec)
Y: >40.6 cm/sec (>16 in./sec)

Communication Interface

RS-232 or GSIOC; Four inputs (contact closure, TTL, or open-collector), four relay outputs, and one switched +24V DC 1A output

Liquid Handling

Reproducibility:
1.37% CV with 50 µL of water
Accuracy:
99.7% with 50 µL of water

Front Panel

8-character display, START and emergency STOP soft keys

Horizontal Motion Strength

X: 5 kg (11.1 lbs.)
Y: 7 kg (15.6 lbs.)

Injection Carryover

<0.05% with 100-µL syringe and total loop filling method

Injection Reproducibility

1.5% CV

Injection Valve Switching Speed

<100 ms

Injection Volume

500 nL–5 mL

Locator Plate Capacity

Up to five Code 200-Series racks, up to seven Code 20- or 30-Series racks (or a combination of up to five racks of both types), or one Code 500-Series and one Code 200-Series rack

Maximum Syringe Flow Rates for Water

Syringe Size
(µL)

Max. Flow Rate
(mL/min)

100

3.37

250

8.43

500

16.87

1000

33.75

5000

100.00

10,000

100.00

25,000

100.00

Power Requirements

Frequency: 50–60 Hz
Voltage: 100–120 and 200–240V, mains voltage fluctuations not to exceed ±10% of the nominal voltage
Current rating: 2.0A for 100–120V or 1.0A for 220–240V

Probe Positioning Performance

Accuracy: ±0.5 mm in X/Y dimensions, ±1 mm in Z dimension
Repeatability: ±0.25 mm in X/Y/Z dimensions

Probe Rinse

Through a dedicated rinse station for rinsing the inside and outside of the probe; selectable rinse volume and flow rate. Optional inserts for level-sensing, nonleveling-sensing and flow rinse

Sampler Type

X/Y/Z with stationary rack design

Software

Via Gilson UniPoint™ System Software or TRILUTION™ Software

Syringe Pump

Integral, high-precision, single-piston

Syringe Capacity

100, 250 or 500 µL; 1 or 5 mL

System Resolution

0.1 µL

Vertical Punch Strength

4.1 kg (11.0 lbs.)

Dimensions

91.4 x 61 x 55.8 cm* (38.5 x 24 x 22* in.)
*Maximum height. Z-arm height is adjustable to accommodate vessel heights between 1 and 150 mm (dependent on installed Z-arm).

Weight

39.9 kg (89 lbs.)