HomeTo Order
 

PipettesPipettorsPipette TipsAccessoriesAutomated Liquid Handlers

InjectorsFraction CollectorsPumps & Tubing

   
215 SW Liquid Handler/Injector

The 215 SW Liquid Handler delivers the same "workhorse" performance as Gilson's 215 Liquid Handler with a smaller footprint to save on bench space, which allows the 215 SW to fit inside a fume hood for safer use while working with volatile chemicals. The increased clearance between the Z-arm and the bed enables the use of larger tubes and bottles.

Features:

  • Accommodates virtually all of Gilson’s numerous standard and custom racks (more than 330), providing increased application flexibility
  • The 215 SW Controller provides complete remote-access control of the instrument from outside a fume hood
  • The external 402 Dual-Syringe Pump increases liquid handling and solvent delivery flexibility; large delivery volume range of 1 µL to 25 mL per piston stroke
  • Optional High-Force Z-Arm provides more than twice the vertical punch strength than the standard Z-arm for piercing thick septa
  • Optional 819 Injection Module enables automated on-line injection onto HPLC systems

Specifications:

Manufacturing Standards

Meets applicable safety and EMC certification standards; UL and CE certified

Arm Speed

X: >45.7 cm/sec (>18 in./sec)
Y: >40.6 cm/sec (>16 in./sec)

Communication Interface
(215 SW Controller)

RS-232 or GSIOC; Four inputs (contact closure, TTL, or open-collector), four relay outputs and one switched +24V DC 1A output

Cross Contamination†

0.001% into first methanol blank

Injection Volume

1 µL–25 mL

Horizontal Motion Strength

X: 4 kg (9 lbs.)
Y: 3 kg (7 lbs.)

Injection Carryover†

0.1%

Reproducibility†

Liquid Handling:
0.8% at 10 µL
Injection:
<0.9% CV with 1-mL syringe and total loop filling method

Locator Plate Capacity

Up to four Code 200-Series racks, four Code 20- or 30-Series racks (or a combination of up to four racks of both types), or one Code 500-Series rack

Power Requirements
(215 SW Controller)

Frequency: 50–60 Hz
Voltage: 100–120 and 200–240V, mains voltage fluctuations not to exceed ±10% of the nominal voltage
Current rating: 3.0A for 100–120V or 1.5A for 220–240V

Probe Positioning Performance

Accuracy: ±0.5 mm in X/Y dimensions, ±1 mm in Z dimension
Repeatability: ±0.25 mm in X/Y/Z dimensions

Probe Rinse

Through a dedicated rinse station for rinsing the inside and outside of the probe

Sampler Type

X/Y/Z with stationary rack design

Software

Via Gilson UniPoint™ System Software or TRILUTION™ Software

Syringe Capacity

100, 250, or 500 µL; 1, 5, 10, or 25 mL

Vertical Punch Strength

5 kg (12 lbs.)
High-Force Z-Arm: 11 kg (25 lbs.)

Accuracy†

Liquid Handling:
98.2% at 10 µL

Dimensions (w x d x h)

78.7 x 61 x 74.9* cm (31 x 24 x 29.5* in.)
*Maximum height. Z-arm height is adjustable to accommodate vessel heights between 1 and 180 mm (dependent on installed Z-arm).
215 SW Controller: 26.2 x 40.6 x 15.2 cm (10.3 x 16 x 6 in.)

Weight

38.6 kg (85 lbs.)
215 SW Controller: 5.9 kg (13 lbs.)