HomeTo Order
 

PipettesPipettorsPipette TipsAccessoriesAutomated Liquid Handlers

InjectorsFraction CollectorsPumps & Tubing

   
Constellation®1200 Liquid Handler

Precise, nanoliter-volume liquid handling capability of the Constellation 1200 Liquid Handler allows you to accurately dispense volumes from 5 nL to 500 µL without an air gap. Precision robotics and flexible software ensure accurate, repeatable X/Y motion so you’ll hit your target every time. Capable of dispensing with or without an air gap, providing increased flexibility for your specific application.

Cat. No. GA-281010

Features:

  • Large bed holds up to 12 microplates
  • Aspirates from 96- and 384-well microplates; dispenses into 96-, 384-, 1,536- and 3,456-well microplates, and MALDI-TOF plates, microscope slides, membranes and other non-standard formats
  • 12-tip Z-arm configuration provides high-throughput liquid handling for genomic and proteomic applications
  • Non-contact dispensing eliminates damage to tips and spotting surface while reducing potential contamination
  • Optional microplate chiller cools plates to reduce sample evaporation and preserve sample integrity

Applications:

  • Protein crystallography
  • Nanoliter serial dilutions
  • MALDI spotting
  • HTS library distribution

Specifications:

Manufacturing Standards

Meets applicable safety and EMC certification standards; UL and CE certified

Bed Capacity

Twelve 96-well, 384-well or 1,536-well shallow microplates

Communication Interface

RS-232 or GSIOC; Four inputs (contact closure, TTL, or open-collector), four relay outputs and two switched +24V DC 1A outputs

Liquid Handling

Reproducibility:
<10% at 50 nL–1 µL; <15% at 5 nL–50nL without an air gap
Accuracy:
±10% at 5 nL–50nL without an air gap

Front Panel

24-character, 2-line LED display and emergency stop activated by opening or pressing the front door panel

Gas Requirement

Helium that is 99.99999% pure; 100–300 ft.3 at 3.4 bar (20 psi) using a gas pressure regulator

Maximum Flow Rates
(for degassed, HPLC-grade water through the probe)

Syringe Size
(µL)

Aspirate Rate
(mL/min)

Dispense Rate
(mL/min)

100

2

2

500 (default)

2

2

1000

2

3

2500

2

3

5000

2

3

Power Requirements

Frequency: 50–60 Hz
Voltage: 100–120 or 220–240V, mains voltage fluctuations not to exceed ±10% of the nominal voltage
Current rating: 3.5A for 100–120V or 1.75A for 220–240V

Probe Positioning Performance

Accuracy: ±0.030 mm in X/Y dimensions over 600 mm
Repeatability: ±0.002 mm in X/Y dimensions; ±0.001 mm in Z dimension

Probe Rinse

Through a dedicated rinse station for rinsing the inside and outside of the probes; selectable rinse time and optional flowing rinse

Sampler Type

X/Y/Z robot with a 12-probe configuration

Software

Via Gilson 735 Sampler Software

Syringe Capacity

100 or 500 µL; 1, 2.5 or 5 mL

Syringe Pump

Dual, internal, integral, high-precision

Dimensions (w x d x h)

Overall: 114.6 x 68.7 x 76.4 cm (45.1 x 27.1 x 30.1 in.)
Base Only (for use inside a fume hood): 93.9 x 67.3 x 45.7 cm (37 x 26.5 x 18 in.)

Weight

120.7 kg (266 lbs.)

  Privacy Policy | Comments