HomeTo Order
 

PipettesPipettorsPipette TipsAccessoriesAutomated Liquid Handlers

InjectorsFraction CollectorsPumps & Tubing

   
GX-281 Liquid Handler – for Injection, Fraction Collection and Re-injection

The new GX-281 Liquid Handler features a revolutionary solvent delivery system: there are no syringes to replace or change to accommodate various volumes and flow rates. The Solvent Delivery System features an innovative technology that allows for a large dynamic range of 5 µL to 100 mL and flow rates of 1 µL/min to 25 mL/min.

Cat. No. GA-2603601


Features:

  • Revolutionary new Solvent Delivery System with integrated solvent selection of up to 5 solvents: No syringes to replace or change to accommodate different volumes and flow rates; allows for immediate access to up to 5 solvents and provides a large dynamic volume and flow rate range
  • Continuous flow path design on all injection valves for flow rates up to 200 mL/min: Allows system flow to continue even as the injection valve switches from load to inject, ensuring a more consistent pressure and flow
  • Direct Injection Module features an injection port that is integrated into the valve: Removes a source of potential carryover; reduces rinsing times, sample waste and X/Y travel for quicker injection cycle times
  • Enhanced bed capacity over other similar liquid handlers—uses only 33" of linear bench space: Holds up to six Code 200-Series racks with an efficient bed design
  • Separate injection and collection pathways: Ensures that fractions are never contaminated from the injection process
  • New rinse station system includes a high-flow rinse pump: Provides fast and efficient probe cleaning with up to 6 rinse solvents

Specifications:

Manufacturing Standards

Meets applicable safety and EMC certification standards; UL and CE certified

Arm Speed

X and Y: 0.01-600.00 mm/sec
Z: 0.01-125.00 mm/sec

Communication Interface

RS-232 or GSIOC; Four inputs (contact closure, TTL), four relay outputs, four switched +24V DC 1A outputs, one safety input and one analog input

Liquid Handling

Reproducibility: <2%
Accuracy: ±2% for (100 µL - 25 mL) water

Front Panel

ON/OFF and STOP soft keys, LED indicator for power

Horizontal Motion Strength

X: 1.4 kg (3 lbs.)
Y: 0.7 kg (1.5 lbs.)
Z: 5.4 kg (12 lbs.)

Injection Carryover

<0.005%

Injection Reproducibility

<1.0% CV (total loop injection), <3.5% (partial loop injection)

Reservoir Solvents

Up to 5

Rinse Solvents

Up to 6

Rinse Speed

High: 200 rpm
Low: 120 rpm

Bed Capacity

Holds up to six Code 200-Series racks
Quantity 576 of 13 mm test tubes
Quantity 450 of 16 mm test tubes
Quantity 420 of 18 mm test tubes
Quantity 180 of 25 mm test tubes
Quantity 62 of 25 or 50 mL scintillation vials

Power Requirements

Frequency: 50-60 Hz
Voltage: 100-240V (Universal Input)
Current rating: 5.0A for 100-120V or 2.5A for 220-240V
Power consumption: 500W maximum

Probe Positioning Performance

Accuracy: ±0.5 mm in X/Y/Z dimensions
Repeatability: ±0.025 mm in X/Y/Z dimensions

Fraction Collector 3-way Valve

3-port, PTFE, 114 µL internal volume, 10 µL dead volume, up to 200 mL/min

Sampler Type

X/Y/Z with stationary rack design

Software

Gilson TRILUTION™ Software

Solvent Delivery Dynamic Range *

2 µL to >100 mL
*For volumes below 100 µL, Liquid Handling and Injection specifications may differ from those listed above.

Flow Rate

Up to 25 mL/min

Liquid Contact Materials

Valcon H, Nitronic 60 (N60), PTFE

Vertical Punch Strength

4.5 kg (10 lbs.)

Display Panel

Two lines of 40 alphanumeric characters, vacuum fluorescent display

Weight

68 kg (150 lbs.)

Dimensions

88.9 x 76.2 x 96.5 cm (35 x 30 x 38 in.)

  Privacy Policy | Comments