Injection |
Mode: total loop or centered loop (except for 50-nL injection loop)
Carryover (with 100-nL internal loop): <0.055%, using rinsing parameters
Reproducibility (with 100-nL internal loop):
• total loop RSD: <2%
• centered loop RSD: <10%
Accuracy: ±5%
Valve switching speed: <0.2 seconds
Volume: 50 nL–100 µL with the standard injection valve (100-nL internal loop), 50 nL–500 nL with loop exchange (4-port valve), 2 µL–100 µL with loop exchange (6-port valve)
Available sample loops:
• internal: 50, 100, 200 or 500 nL
• external: 2, 5, 10, 20, 50 or 100 µL
Transfer flow rate: 2 µL/min (recommended); adjustable from 100 nL/min–100 µL/min |
Sampler Type |
Mobile sample platform cartesian robot |
Maximum Number of Sample Vessels/Bed |
For samples: 2 microplates (96- and 384-well, deep or shallow, or 1,536-well) or 112 x 2-mL vials; For reagents: 4 x 20-mL or 4 x 2-mL vial |
Probes(L x OD x ID) |
Probe options:
• sample prep (stationary): beveled-tip, stainless steel, grooved septum-piercing, constricted (103 x 1.5 x 0.4 mm)
• sample injection (stationary): beveled-tip, stainless steel, grooved septum-piercing, constricted (103 x 1.5 x 0.4 mm; tip OD: 0.7 mm) with fused silica tubing (263 x 0.36 x 0.1 mm)
• sample injection (stationary): flat tip for 1,536-well plates, stainless steel (103 x 2 x 0.4 mm; tip: 6 x 0.36 x 0.1 mm) with fused silica tubing (263 x 0.36 x 0.1 mm) |
Probe Rinse |
Inside probe rinse via dedicated syringe pump; outside probe rinse via peristaltic pump |
Tray Positioning Performance |
Accuracy: ± 0.05 mm (0.002 in.) in X/Y/Z dimensions |
Tray Speed Performance |
75 mm/sec. (3 in./sec) in X/Y dimensions; 50 mm/sec. (2 in./sec) in Z dimension |
Tray Cooling |
Sample cooling via external chiller |
Syringe Pumps |
Three syringe pumps for: high-precision injection, injection probe rinsing and sample preparation |
Reccommended Syringe Capacities |
Injection syringe: 10, 50, 100 or 500 µL
Rinsing syringe: 100, 500, 1000 or 2500 µL
Sample prep. syringe: 100, 500, 1000 or 2500 µL |
Available Valves |
Standard: one 4-port valve; optional: two additional 6-port valves |
Software |
Gilson 735 Sampler Software |
Communication Interface |
RS-232 or GSIOC; Three relay outputs, one 12V DC, 0.5A output, three contact inputs, and one analog signal input |
Front Panel |
Power, Detector Calibration, and Air Gap LED indicators; Detector, Dilutors, HP Valves, and Arm ERROR LED indicators; emergency STOP soft key |
Liquid Contact Materials |
PEEK, Teflon®, 316 stainless steel, fused silica |
Power Requirements |
Current rating: 3.15A
Frequency: 50–60 Hz
Voltage: 100–120V or 220–240V, mains voltage fluctuations not to exceed ±10% of the nominal voltage |
Environmental Operating Temperature |
5–40°C |
Manufacturing Standards |
Meets applicable Safety and EMC certification standards; UL and CE certified |
Dimensions (w x d x h) |
Sample platform: 74 x 54 x 39 cm (29 x 21 x 15 in.)
Hydraulic module: 41 x 25 x 23 cm (16 x 10 x 9 in.) |
Instrument Weight |
Sample platform: 47 kg (104 lbs.)
Hydraulic module: 10 kg (22 lbs.) |
Shipping Weight |
75 kg (165 lbs.) |