YSI TruLab 1310 pH/mV single channel benchtop instrument with integrated thermal printer. AutoRead function improves repeatability, Easy-to-read large, segmented display for pH and mV (ORP); easy calibration with adjustable calibration timer; automatic or

The Largest 100% Canadian Owned and Operated Lab Equipment Distributor and Service Provider.

Free Technical Support
In-depth and personal over-the-phone support at no cost.

Quality Parts & Accessories
Mandel uses manufacturer-specified parts for all repairs and maintenance.
Details
The YSI TruLab line includes the 1110 (single channel), 1310 and 1310P (single channel) and 1320 and 1320P (dual channel) instruments providing easy-to-use and calibrate instruments ideal for the laboratory.
The TruLab pH 1310 benchtop meter is the ideal laboratory instrument for reliable pH/mV (ORP) measurements with automatic documentation for GLP compliance.
Available with or without a built-in printer. The 1310 has internal memory. Stored data can be sent to a PC via USB or to an optional built-in printer available on the 1310P.
Features
Features
- Single channel measurement of pH or mV (ORP) and temperature
- AutoRead function improves repeatability
- Easy-to-read backlit display with text menu for easy operation for pH, mV (ORP) and temperature
- Fast data transfer via USB in .csv or ASCII format or the optional integrated printer
- Easy calibration with adjustable calibration timer
- Automatic or manual temperature compensation
- 1, 2, 3, 4 or 5 point calibration for pH; 22 stored buffer sets for auto-buffer recognition and easy calibration
- BNC connection for pH, mV; banana connection for temperature probes
- Continuous Monitoring Control (CMC) function for monitoring the ideal measuring range for pH measurements based on the last calibration
- 3-year warranty
- GLP/AQA compliant
- PC software, USB cable, and electrode stand included
- Wide selection of electrodes - pH, ORP and temperature probes
Repeatable/Efficient
The instrument's AutoRead function indicates when a stable measurement is reached improving repeatability.
Convenient/Reliable
Data can be printed directly from the instrument via the integrated printer or exported via USB.
Specifications
Benchtop | Yes |
---|---|
Buffer Recognition | Yes |
Certifications | cETLus, IP43, electrical protective class III, CE |
Display | Backlit graphic display |
Equipment used with | TruLab analogue pH/mV/temp probes |
GLP Compliance | Yes |
Graphic Display | Backlit graphic display |
Logging Capabilities | Yes |
Measurement Method / Channels | 1 Channel |
Measurement Range | Ranges Resolutions Accuracy (+1 digit) pH -2.0 to 20.0; -2.00 to 20.00; -2.000 to 19.999 0.1; 0.01; 0.001 ±0.1; ±0.01; ±0.005 (temp of test sample 15 to 35°C; within a range of ±2 pH of the calibration point) mV -1200 to 1200.0; -2000 to 2000 0.1; 1.0 ±0.3; ±1.0 (temp of test sample 15 to 35°C) Temperature -5 to 105°C (23.0 to 221°F) 0.1 ±0.1 |
Memory | 500 data sets manual logging, 5,000 data sets auto logging |
Operating pH Range | -2.00 to 20.00, -2.0 to 20.0, -2.000 to 19.999 |
Operating Temperature | 23.0 to 221°F, -5 to 105°C |
Parameters Measured | pH, ORP (mV), Temperature |
Power | 4 alkaline AA batteries (~150 hours with illumination, 1000 without), compatible with 4 NiMH AA rechargeables. Instrument does not charge rechargeable batteries., universal power supply |
Sampling | Yes |
Storage Temperature | -25 to 65°C, -13 to 149°F |
User Calibratable | Yes |
User Replaceable | Yes |
Warranty | 3 years |
- Petri Dish, Quad Plate, Stackable,100 x 15, Case/500 in sleeves of 25Regular Price CA$238.00 Special Price CA$214.20