Nexera HPLC/UHPLC Autosamplers

Mandel brings you a range of impressive Shimadzu UHPLC and HPLC autosamplers for your HPLC instrumentation.


HPLC/UHPLC Autosampler - SIL-40C X3

For highest efficiency UHPLC with sub-2 micron columns. A new benchmark in speed and low carryover.

Multi-rinsing is standard.

Allowable maximum pressure 130 MPa
Injection volume 0.1 to 50 μL (0.1 to 2000 μL - optional)
Carryover

≤ 0.0015% (without rinse)
≤ 0.0003% (with rinse, typically)
(under specified conditions)

Pumping rinse outside the needle Standard equipment
Internal rinsing (3 dil) Standard equipment
Sample cooler

Standard equipment
(Air-circulation temperature control type)

Weight 24 kg


Nexera HPLC/UHPLC Autosampler - SIL-40C XS

For moderate efficiency UHPLC with sub-2 micron and SPP columns. A new benchmark in speed and low carryover.

Multi-rinsing is optional.

Allowable maximum pressure 105 MPa
Injection volume 0.1 to 50 μL (0.1 to 2000 μL - optional)
Carryover

≤ 0.0015% (without rinse)
≤ 0.0003% (with rinse, typically)
(under specified conditions)

Pumping rinse outside the needle Standard equipment
Internal rinsing (3 dil) Option
Sample cooler

Standard equipment
(Air-circulation temperature control type)

Weight 24 kg

Nexera HPLC/UHPLC Autosampler - SIL-40/SIL-40C

For HPLC with SPP and conventional columns. Like the UHPLC models, a benchmark in speed and low carryover.

Multi-rinsing is optional.

Allowable maximum pressure 44 MPa
Injection volume 0.1 to 100 μL (0.1 to 2000 μL - optional)
Carryover

≤ 0.0025% (without rinse)
≤ 0.0005%
(with rinse, typically)
(under specified conditions)

Pumping rinse outside the needle Option
Internal rinsing (3 dil) Option
Sample cooler

SIL-40: None
SIL-40C:
Standard equipment
(Air-circulation temperature
control type)

Weight

SIL-40: 17 kg
SIL-40C: 24 kg

Power supply (non-cooler model)

AC 100–240 V,150 VA, 50/60 Hz


HPLC/UHPLC Plate Changer

Expands the capacity of any 40-Series autosampler by up to 14 vial racks or MTPs. Up to three plate changers may be used on one system.

Samples for processing:
1 Plate Changer

(Includes two plates of autosampler)

1536 (microtiter plate, 96 well × 16 plates), 864 (deep-well plate, 96 well × 9 plates)
6144 (microtiter plate, 384 well × 16 plates), 3456 (deep-well plate, 384 well × 9 plates)
756 (1 mL sample vial, 84 × 9 plates), 486 (1.5 mL sample vial, 54 × 9 plates)
252 (4 mL sample vial, 28 × 9 plates), 108 (10 mL sample vial, 12 × 9 plates)

Samples for processing:
3 Plate Changers

(Includes two plates of autosampler)

4224 (microtiter plate, 96 well × 44 plates), 2208 (deep-well plate, 96 well × 23 plates)
16896 (microtiter plate, 384 well × 44 plates), 8832 (deep-well plate, 384 well × 23 plates)
1932 (1 mL sample vial, 84 × 23 plates), 1242 (1.5 mL sample vial, 54 × 23 plates)
644 (4 mL sample vial, 28 × 23 plates), 276 (10 mL sample vial, 12 × 23 plates)

Sample cooler
temperature setting range

Air-circulation temperature control type, 4 to 45°C
(Room temperature needs to be less than 30°C and humidity needs to be less than 70% to set 4°C)

Dimensions [mm], weight W 170 × D 500 × H 560 (Protrusion adds 140 mm to the depth), 26 kg
Operating temperature range 4 to 35°C
Power supply

AC 100−240 V, 400 VA, 50/60 Hz


Universal Specifications

Injection method Total-volume Injection (standard), loop injection (optional)
Injection volume accuracy ≤ ± 1% (5 μL injection, n = 20)
Linearity ≥ 0.9999
Injection cycle time ≤ 6.7 seconds (under specified conditions)
Samples for processing

288 (microtiter plate, 96 well × 3 plates), 1152 (microtiter plate, 384 well × 3 plates),
252 (1 mL sample vial, 84 × 3 plates), 162 (1.5 mL sample vial, 54 × 3 plates),
84 (4 mL sample vial, 28 × 3 plates), 36 (10 mL sample vial, 12 × 3 plates),
72 (1.5 mL micro tube, 24 × 3 plates)

Injection volume reproducibility

RSD ≤ 1.0% (0.5 to 0.9 μL), RSD ≤ 0.5% (1.0 to 1.9 μL), RSD ≤ 0.25% (2.0 to 4.9 μL),
RSD ≤ 0.15% (More than 5.0 μL), RSD < 0.5% (typically, 0.5 μL), RSD < 0.25% (typically, 1.0 μL)

Dip rinsing outside the needle
and injection port rinsing

Standard equipment

Sample cooler
temperature setting range

4 to 45°C
(Room temperature needs to be less than 30°C and humidity needs to be less than 70% to set 4°C)

Sample cooler
temperature accuracy

± 2°C (sensor position ± 0.5°C)
Wetted material SUS316L, DLC, PEEK, GFP, PTFE, FEP, ETFE, sapphire, ceramics, PPS, FFKM
Available pH range 1 to 14
Dimensions [mm] W 260 × D 500 × H 280 (SIL-40C/40C XR/40C XS/40C X3: Protrusion adds 140 mm to the depth)
Operating temperature range 4 to 35°C
Power supply (cooler model) AC 100–240 V, 400 VA, 50/60 Hz